7HL Tecnosens
- Manufacturer: Tecnosens
MEASUREMENT Operating Principle: 3-electrode electrochemical Measurement Range: 0-50 ppm HCl Maximum overload: 100 ppm HCl Filter: None Sensitivity: 0.75 ± 0.25 μA/ppm Response Time (T90): <120 Seconds at 20°C Ze...
Brescia, Italy
5SFF Tecnosens
- Manufacturer: Tecnosens
MEASUREMENT Operating Principle: 3-electrode electrochemical Nominal Range: 0-2000 ppm SO2 Maximum Overload: 5000 ppm SO2 Filter: To remove effects of H2S and HCl Filter Life: >200,000 ppm hours (1000 ppm H2S @ 5...
Brescia, Italy3SFF Tecnosens
- Manufacturer: Tecnosens
MEASUREMENT Operating Principle: 3-electrode electrochemical Nominal Range: 0-2000 ppm SO2 Maximum Overload: 5000 ppm SO2 Filter: To remove H2S & HCl Sensitivity: 0.10 ± 0.02 μA/ppm Response Time (T90): <35 Secon...
Brescia, Italy
7HLM Tecnosens
- Manufacturer: Tecnosens
MEASUREMENT Operating Principle: 3-electrode electrochemical Measurement Range: 0-200 ppm H2S Maximum overload: 1000 ppm H2S Filter: None Sensitivity: 0.37 ± 0.07 μA/ppm Response Time (T90): <35 Seconds at 20°C Z...
Brescia, Italy
7HH Tecnosens
- Manufacturer: Tecnosens
MEASUREMENT Operating Principle: 3-electrode electrochemical Measurement Range: 0-50 ppm H2S Maximum Overload: 500 ppm H2S Filter: None Sensitivity: 1.7 ± 0.3 μA/ppm Response Time (T90): <30 Seconds at 20°C Basel...
Brescia, Italy
7HHLM Tecnosens
- Manufacturer: Tecnosens
MEASUREMENT Operating Principle: 3-electrode electrochemical Measurement Range: 0-50 ppm H2S Maximum overload: 500 ppm H2S Filter: None Sensitivity: 1.7 ± 0.3 μA/ppm Response Time (T90): <30 Seconds at 20°C Zero ...
Brescia, Italy
3HH Tecnosens
- Manufacturer: Tecnosens
MEASUREMENT Operating Principle: 3-electrode electrochemical Measurement Range: 0-50 ppm H2S Maximum Overload: 500 ppm H2S Filter: None Sensitivity: 1.70 ± 0.30 μA/ppm Response Time (T90): <30 Seconds Baseline Of...
Brescia, Italy
7ETO Tecnosens
- Manufacturer: Tecnosens
MEASUREMENT Operating Principle: 3-electrode electrochemical Measurement Range: 0-20 ppm C2H4O Maximum Overload: 100 ppm C2H4O Filter: None Sensitivity: 2.25 ± 0.65 μA/ppm Response Time (T90): <140 Seconds Baseli...
Brescia, Italy
TGS8410 Tecnosens
- Manufacturer: Tecnosens
The sensing element consists of a metal oxide semiconductor layer formed on a silicon substrate of a chip with an integrated heater. In the presence of the gas to be sensed, the conductivity of the sensor increas...
Brescia, Italy
4ETO Tecnosens
- Manufacturer: Tecnosens
MEASUREMENT Operating Principle: 3-electrode electrochemical Measurement Range: 0-20 ppm C2H4O Maximum Overload: 100 ppm C2H4O Resolution (Electronics dependent): 0.1 ppm Filter: None Sensitivity: 1.9 ± 0.5 mA/pp...
Brescia, Italy
7OZ Tecnosens
- Manufacturer: Tecnosens
MEASUREMENT Operating Principle: 3-electrode electrochemical Measurement Range: 0-2ppm O3 Maximum Overload: 5 ppm O3 Filter: None Sensitivity: 7.2 mA/ppm ±2.3 mA/ppm Response Time (T90): <150 seconds @ 20°C Basel...
Brescia, Italy