Microtrac SYNC Combining Laser Diffraction with Dynamic Image Analysis in Worthington, OH, USA

Specifications

Condition
new
Measuring range
0.01 µm - 4 mm
Measuring principle
Laser Diffraction (ISO 13320)Dynamic Image Analysis (ISO 13322-2)
Precision*
Spherical glass beads D50 = 642 microns, precision as CV = 0.7% Spherical glass beads D50 = 57 microns, precision as CV = 1.0% Spherical latex beads D50 = 0.4 microns, precision as CV = 0.6%
Laser class
Red 780 nm, blue 405 nm Class 1 laser product per CFR 1040.10 & IEC60825-1
Laser power
Red laser 0.35 to 2 mW nominal Blue laser 4-8 mW nominal
Detection system
Two fixed photo-electric detectors with logarithmically spaced segments are placed at correct angles for optimal scattered light detection from 0 to 165 degrees using 151 detector segments.
Data
Volume, number and area distributions as well as percentile and other summary data
Data format
Stored in ODBC format in encrypted Microsoft Access Databases to ensure compatibility with external statistical software applications.
Data integrity
Data integrity may be ensured using FDA 21 CFR Part 11 compliant security features including password protection, electronic signatures and assignable permissions
Type of analysis
dry and wet analysis
Measuring time
~ 10 to 30 seconds
Power requirements
AC input: 90 - 264 VAC, 47 - 63 Hz, single phase
Power consumption
25 W nominal, 50 W max., depending on options installed
Environmental conditions
Temperature: 5° to 40° Celsius (50° to 95° Fahrenheit) Humidity: 90% RH, non-condensing maximum Storage Temperature: -10° to 50° Celsius (14° to 122° Fahrenheit) (dry only) Pollution: Degree 2
Standards
Laser Diffraction (ISO 13320)Dynamic Image Analysis (ISO 13322-2)Representation of results of particle size analysis (ISO 9276-6)
Image analysis
5.2 megapixel (2560 x 2048), 22 fps at max resolution
Wet operation
Volume: 200 ml nominal Flow rate: 0 to 65 ml/sec with water Inlet pressure: 50 psig (345 kPa) maximum
Dry operation
100 psi (689 kPa) maximum pressure 5 CFM (8.5 m3/h) at 50 psi (345 kPa) minimum flow rate Free of dry contaminants, moisture and oil
Vacuum
Vacuum must meet or exceed 50 CFM (85 m3/h)
Physical specifications
Case Material: Impact resistant plastic Exterior surfaces are finished with corrosion resistant paint or plating Chemical Compatibility: Class I
Dimensions (w x h x d)
~ 820 x 460 x 500 mm (32.3 x 18.1 x 19.7 in)
Weight (measuring unit)
FlowSync: 19.5kg (43lbs)TurboSync: 13.6kg (30lbs)Sync: 23.6kg (50.8lbs)
Subcategory
Sync
Subcategory 2
Laser diffraction
Subcategory 3
Particle size & shape analysis
Listing ID
92640612

Description

A New Dimension in Particle Analysis
With the SYNC particle analyzer, Microtrac integrates its highly accurate tri-laser diffraction analyzer technology with its versatile dynamic image analysis capability to provide particle characterization practitioners with a unique measuring experience. The patented synchronous measurement technology allows users to make both a laser diffraction measurement and an image analysis measurement on a single sample, in the same sample cell at the same time:
One sample
One optical bench
One flow path

One sample cell
One analysis
The SYNC laser diffraction particle size analyzer is ideally suited for routine QC applications. It also provides valuable information to researchers as they develop new materials and processes. The powerful analyzer software provides both particle size distribution information as well as a multitude of morphological particle parameters.

Contact Seller for Price

Manufacturer
Microtrac
Location
🇺🇸 Worthington, OH, USA

Interested in this machine?

This seller has been contacted 1 time in the last week.

Seller Responsiveness:

4.75
Get Financing

Our financing partner can help you get fast, easy, and affordable financing for your equipment purchase