Dionex Ics-5000 Ion Chromatography System in Markham, Ontario, Canada
Refurbished
Doubleclick to zoom in
Contact the seller for
additional photos and information.
Specifications
- Condition
- refurbished
- Stock number
- Dionex-02-0630
- Pressure range
- 0–41 MPa (0–6000 psi)
- Flow rate range
- Analytical: 0.000–10.000 mL/min with settable flow increments at 0.001 mL/min Capillary: 0.001–3.000 mL/min with settable flow increments at 0.0001 mL/min
- Flow rate precision
- <0.1%
- Flow rate accuracy
- <0.1% at 14 MPa (2000 psi)
- Pressure ripple
- Analytical: < 1.0% at 1.0 mL/min typical Capillary: < 0.2% (with damper) or < 1.0% (without damper) at 10 μL/min
- Pressure
- None required
- Vacuum degasser
- Integrated degasser with 1 channel for isocratic pump or 4 channels for quaternary pump; optional
- Piston seal wash
- Automatic operation, standard
- Gradient formation
- Quaternary low pressure (Analytical only) or electrolytic eluent generation at high pressure (Analytical and Capillary)
- Rfic-eg gradients
- Analytical: 0.1–100 mM Capillary: 0.1–200 mM
- Gradient profiles
- Standard—any combination of an unlimited number of linear, convex, and concave positive and negative gradient profiles
- Gradient proportioning accuracy and precision (analytical only)
- ±0.5% at 2 mL/min
- Gradient mixing (analytical only)
- Passive mixers for 2 mm and 4 mm ID columns, Optional
- Eluent generation
- Eluent generation (RFIC-EG), optional
- Eluent on/off valve
- Electrically-actuated, standard
- Leak sensor
- Optical, standard
- Minimum and maximum eluent concentrations
- 0.01–100 mM (Analytical) 0.01–200 mM (Capillary)
- Flow rates
- 0.100–3.000 mL/min (Analytical) 0.001–0.030 mL/min (Capillary)
- Eluent types
- KOH, LiOH, NaOH Carbonate Carbonate/bicarbonate Carbonate with pH modifier MSA
- Maximum operating pressure
- 35 MPa (5000 psi)
- Maximum solvent concentration
- Cations: None Anions: 25% methanol
- Number of cartridges supported
- Two—dual cartridge support
- Dc models:
- Dual Temperature Model Only
- Dual temperature model: independent temperature control
- Upper Section Temp. Range: 10–40 °C Lower Section Temp. Range: 10–70 °C Temperature Accuracy: ± 0.5 °C Temperature Stability: ± 0.2 °C Temperature Precision: ± 0.2 °C
- Lower zone: (analytical)
- Injection valves: 1 or 2 ea., 6 or 10 port Up to two column sets, 1–9 mm Maximum column length: 250 mm Precolumn heat exchangers: 2
- Cap ic cube: (optional – but required for capillary chemistries)
- 1 or 2 IC Cube™ modules hold the following: Injection Valves: Up to 2 (one per cube) high pressure valves, 4-port, 2 position EG Degasser Column and Guard Carbonate Removal Device Capillary Suppressor (See Suppressor section) IC Cubes and cartridges are customer installable with preformed tubing and color-coded labeling Capillary column heater 15 to 85 °C
- Automation manager: (optional)
- Injection valves: Up to 2 ea., high pressure valves, either 6 or 10 port Low pressure valves: Up to 2 ea., PTFE, 2 or 3 way Reaction Coil Heater: Holds up to two reaction coils Temperature Range: 5 °C above DC, up to 80 °C
- Leak detection:
- Optical leak sensor
- Injections per vial
- 1 to 99
- Vial sizes available
- 0.3 mL polymer vials (P/N 055428, package of 100) 1.5 mL glass vials (P/N 055427, package of 100) 10 mL PolyVials (P/N 055058, package of 100)
- Variable volume range
- 1 to 100 µL in 0.1 µL increments 100 to 1000 mL in 1 µL increments
- Injection loop size
- IC version: 25 µL standard; other sizes are available HPLC version: 20 µL standard; other sizes are available
- Injection
- Two-position, six-port, electrically-activated Rheodyne valve IC version: PEEK with Tefzel® rotor seal HPLC version: Stainless steel with Vespel® rotor seal
- Column switching
- Two-position, ten-port, electrically-activated Rheodyne valve IC version: PEEK with Tefzel rotor seal HPLC version: Stainless steel with Vespel rotor seal
- Category
- Chromatography in Canada
- Subcategory
- Thermofisher hplc
- Subcategory 2
- Hplc, uplc & ic products
- Subcategory 3
- Thermo ion chromatography systems
- Listing ID
- 26189195
Description
Features
The ICS-5000 Reagent-Free™ IC (RFIC™) system with Eluent Generation is the world’s first capillary IC system. Developed with flexibility, modularity, and ease-of-use in mind, the ICS-5000 combines the highest sensitivity with convenience.
With the ability to analyze samples at capillary, microbore, or standard flow rates (or any combination of two, in a dual system), the system is simply Always Ready. The wide variety of ICS-5000 modules allows you to configure an ion chromatography system designed specifically to provide the solutions you need.
The Dionex ICS-5000 is the first system with an all-PEEK flow path capable of operating continuously at 5000 psi. Reagent-Free HPIC systems redefine the way ion chromatography is performed.
Dionex ICS-5000 Dual Pump (DP)
Dionex ICS-5000 Eluent Generator (EG)
Dionex ICS-5000 Detector/Chromatography Module
Dionex AS-50 Autosampler
A 90-day warranty
Dionex ICS-5000 Dual Pump(DP)
Dionex ICS-5000 Eluent Generator (EG)
Services offered by seller
This Seller offers repair & maintenance services
This seller offers spare parts